ATEK Docs

ELPRO integration

To retrieve the ELPRO PT100 probes, simply connect the DB9-RJ22 adapter compatible with the ATEK RTD transmitter . Since ELPRO probes have a 4-wire configuration, there is no loss of accuracy due to the electrical resistance that the connector may add. The advantage of ELPRO probe recovery is that

To retrieve the ELPRO PT100 probes, simply connect the DB9-RJ22 adapter compatible with the ATEK RTD transmitter .

Since ELPRO probes have a 4-wire configuration, there is no loss of accuracy due to the electrical resistance that the connector may add.

The advantage of ELPRO probe recovery is that it is not necessary to defrost the freezer to replace the probes.

Following rapid integration, all you need to do is perform a calibration.